EC/TDS/Resistivity/Salinity Meter
The HI98192 is supplied with the HI763133 four-ring conductivity probe with a built in temperature sensor. This technology allows for accurate readings that cover a complete range with a single sensor. The four ring technology also eliminates the polarization effect that is common with standard two pole versions. The HI763133 conductivity probe connects to the meter with a unique quick connect waterproof DIN connector allowing for a secure attachment without a threaded connection that is also easy to remove.
HI98192 has an expanded conductivity range from 0.000 ?S/cm to 1000 mS/cm actual conductivity (400 mS/cm temperature compensated). The HI98192 also has ranges for TDS, resistivity and three salinity scales. This meter automatically recognizes the probe type (two or four ring) and allows the user to adjust the nominal cell constant.
The compact, durable HI720192 carry case is thermoformed to hold all necessary components for taking a field measurement, including: the meter, probe, beakers and conductivity standards.
EC Range 0.000 to 9.999 ?S/cm*; 10.00 to 99.99 ?S/cm; 100.0 to 999.9 ?S/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100.0 to 1000.0 mS/cm (actual conductivity; temperature compensated to 400 mS/cm)
EC Resolution 0.001 ?S/cm; 0.01 ?S/cm; 0.1 ?S/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm
EC Accuracy ±1% of reading (±0.01 ?S/cm or 1 digit, whichever is greater)
EC Calibration automatic up to five points with seven memorized standards (0.00 ?S/cm, 84.0 ?S/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm)
TDS Range 0.00 to 99.99 ppm; 100.0 to 999.9 ppm; 1.000 to 9.999 ppt (g/L); 10.00 to 99.99 ppt (g/L); 100.0 to 400.0 ppt (g/L)
TDS Resolution 0.01 ppm; 0.1 ppm; 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L)
TDS Accuracy ±1% of reading (±0.05 mg/L (ppm) or 1 digit, whichever is greater)
Resistivity Range 1.0 to 99.9 ?•cm; 100 to 999 ?•cm; 1.00 to 9.99 K?•cm; 10.0 to 99.9 K?•cm; 100 to 999 K?•cm; 1.00 to 9.99 M?•cm; 10.0 to 100.0 M?•cm
Resistivity Resolution 0.1 ?•cm; 1 ?•cm; 0.01 K?•cm; 0.1 K?•cm; 1 K?•cm; 0.01 M?•cm; 0.1 M?•cm*
Resistivity Accuracy ±1% of reading (±10 ? or 1 digit, whichever is greater)
Salinity Range % NaCl : 0.0 to 400.0%; practical salinity: 0.00 to 42.00 (PSU); natural seawater scale - UNESCO 1966: 0.00 to 80.00 (ppt)
Salinity Resolution 0.1%; 0.01
Salinity Accuracy ±1% of reading
Salinity Calibration max. one point only in % range (with HI7037 standard); use conductivity calibration for all other ranges
Temperature Range -20.0 to 120.0°C (-4.0 to 248.0°F)
Temperature Resolution 0.1°C (0.1°F)
Temperature Accuracy ±0.2°C; ±0.4°F (excluding probe error)
Temperature Calibration one or two points
Temperature Compensation noTC, Linear (-20 to 120 °C; -4 to 248°F), Non Linear - ISO/DIN 7888 (-0 to 36 °C; 32 to 96.8°F)
Measurement Modes autorange, autoend, lock and fixed range
Reference Temperature 15°C, 20°C and 25°C
Temperature Coefficient 0.00 to 10.00 %/°C
TDS Factor 0.40 to 1.00
Electrode/Probe HI763133, four ring conductivity/TDS probe with internal temperature sensor and 4 m (13.2’) cable (included)
Log-on-demand 400 samples; lot logging: 5, 10, 30 sec, 1, 2, 5, 10, 15, 30, 60, 120, 180 min (max 1000 samples)
Memorized Profiles up to 10
PC Connectivity opto-isolated USB with optional HI 92000 software and micro USB cable
Environment | 0 to 50°C (32 to 122°F); RH 100% (IP67) |
Battery Type | 1.5V AA batteries (4) / approximately 100 hours of continuous use (without backlight), 25 hours with backlight |
Auto-off | 5, 10, 30, 60 min or can be disabled |